Documented headlight bench test · 18 Nov 2026

AUXITO H4 LED Headlight Test:
What Did the Lux Readings Show?

An unedited-measurement, edited-for-viewing test: AUXITO H4 bulb inspection and installation, an approximately 3 m setup, separate low/high-beam readings, then a full 4m 34s low-beam Continuous Test played at 10× speed.

Measurements made with Letrava HL200 · Specific sensor positions · Not a total-lumen or road-compliance test
Low-beam mode · Low-beam sensor1,357 lxInitial point-illuminance reading
High-beam mode · High-beam sensor768 lxDifferent sensor position
Continuous low-beam record+4.0%APP-indicated change vs. starting reference
Original testing footage

Watch the complete sequence, with the full timed test accelerated.

English vertical video. The headlight and phone readings are real filmed scenes; the Continuous Test interval is shown from start to finish at 10× playback speed, followed by the APP statistics.

Inside the 69-second edit

  1. AUXITO H4 LED bulb shown and installed in a headlight assembly
  2. Tape-measured test separation of approximately 3 m
  3. Low-beam mode: 1,357 lx at the low-beam sensor position
  4. High-beam mode: 768 lx at the high-beam sensor position
  5. Return to low-beam mode: 1,356 lx recorded
  6. Full 4m 34s Continuous Test at 10× speed, followed by APP results
Open or download the original edited MP4

The 10× interval is an accelerated presentation of the recorded footage. It is not a real-time playback of four and a half minutes.

Test setup

Six steps, one documented measurement setup.

The original clips were reorganized into the actual testing order. The footage shows an AUXITO H4 LED bulb installed in a headlight assembly, with the distance checked by tape before measuring the illuminance reaching the HL200 sensors.

Actual AUXITO H4 LED bulb held during the original test
01 / BULB

Identify the LED bulb

Real footage documents the AUXITO H4 lamp and its physical construction.

Installing the AUXITO H4 LED bulb into the headlight assembly
02 / INSTALLATION

Install into headlight assembly

The bulb is fitted to the test headlight housing before illumination is measured.

Real measuring tape used to check the headlight bench test distance
03 / TEST DISTANCE

Approximately 3 meters

A tape measure records the test separation; the results apply to this setup, not all distances.

Low beam → High beam → Back to low

What the HL200 APP actually displayed.

These are illuminance readings at two fixed sensor positions, recorded under different beam modes. The low-beam and high-beam values are not interchangeable measurements of total bulb output.

HL200 application low-beam mode, low-beam sensor showing approximately 1357 lux
01 · LOW BEAM ON
1,357 lx

Illuminance reaching the low-beam sensor position in low-beam mode.

HL200 application in high-beam mode, high-beam sensor showing approximately 768 lux
02 · HIGH BEAM ON
768 lx

Illuminance reaching the high-beam sensor position after switching modes.

HL200 application after returning to low-beam mode showing approximately 1356 lux
03 · BACK TO LOW BEAM
1,356 lx

Low-beam sensor reading after switching from high beam back to low beam.

Important: 1,357 lx at the low-beam position cannot be compared directly with 768 lx at the high-beam position to conclude that the low beam produces more total light. Lux measures light falling on a specific sensor; lumens describe total luminous flux. Beam geometry, distance, and alignment matter.

Low-beam Continuous Test

4 minutes 34 seconds: illuminance increased at the measured point.

After returning to low-beam mode, the HL200 recorded approximately 4m 34s of continuous data. The entire filmed interval is presented at 10× speed in the video above. The table below reproduces the displayed low-beam sensor statistics from the final APP result screen.

APP measurement fieldLow-beam sensor position
Maximum (MAX)1,463 lx
Minimum (MIN)1,362 lx
Average (AVG)1,426 lx
Current at end1,436 lx
APP-indicated change vs starting reference+4.0%
Recorded Continuous Test video length4m 34s

What the +4.0% does—and does not—mean

The APP indicated an increase of 4.0% at the low-beam sensor position relative to the instrument's starting reference. It does not establish that the LED bulb's total lumen output or electrical power rose by 4.0%.

The 1,463 lx maximum and 1,362 lx minimum reveal a wider observed range during the timed record. This maximum-to-minimum spread is a different statistic from the APP's reference-to-current percentage.

Possible explanations for time-dependent changes would require further controlled trials; this short bench test alone does not identify the underlying cause.

Unaltered frame showing the recorded HL200 Continuous Test graph and final application statistics

Interpretation & limitations

Reading these results correctly.

Why is the low-beam reading higher than the high-beam reading?

They come from different sensor positions under different beam modes. The readings alone do not determine which light distribution produces better road illumination or whether the system is compliant.

Did the LED bulb itself become 4% brighter?

Not established. The APP indicates a 4.0% increase in measured point illuminance relative to its starting reference. It is not a total-lumen or power measurement.

Was the entire Continuous Test shown?

Yes. The submitted 4m 34s continuous-recording segment was played in chronological order at 10× speed in the edited English video. The source segment was not represented as real time.

Does the result prove improved visibility or reduced glare?

No. A single bench setup, two sensor positions and a timed lux record do not establish road performance, glare compliance or long-term durability.